Modes of Operation:
Adult and Pediatric
Volume-Controlled (VCV), Assist/Control
Pressure-Controlled (PCV), Assist/Control Pressure
Support (PSV) CPAP +
PSV SIMV
(VCV) + PSV SIMV
(PCV) + PSV
Mandatory Minute Ventilation (MMV) + PSV PSV +
Tidal Volume Assured
Neonatal
Pressure Controlled (PCV), Assist/Control Pressure
Support (PSV) CPAP SIMV
(PCV) + PSV
Continuous Flow, Time Cycled, Pressure Limited
Apnea Ventilation (back-up): VCV or PCV for
adult and pediatric patients; PCV for
neonates.
Panel Controls and Operating
Range
Tidal Volume: 50 to 2500 mL (Adult and Pediatric
Category) Flow Waveform: In VCV: Square,
Decelerating, Sinusoidal, Accelerating. In PCV and PSV:
Decelerating. Pressure Control (PCV): 1-50 cm H2O, with
rise time. Pressure Support (PSV): 0-50 cm H2O, with
rise time and exp. sensitivity. Rise Time: Modify
the pressure slope. Expiratory Sensitivity: 40, 33,
25 (default), 15, 10 and 5% of peak flow. Respiratory
Rate: 1-150 bpm Sensitivity (PEEP compensated):
Trigger pressure: 0.5 to 10 cm H2O Trigger flow: 0.5 to 5 L/min Inspiratory
Time: 0.1 - 3 seconds PEEP/CPAP: 0 to 50 cm H2O Sigh (VCV only): Number, Events per hour,
Volume, and Max.Pressure. Insp. Pause (VCV only):
0.25 to 2 seconds with 0.25 s increments. I:E Ratio:
Volume mode: 1:1 1:20; Pressure Control: 5:1
1:20 FIO2 : 0.21 1.0 O2 100%:
120 s with 30 s in the middle time for
suction. Stand-by: To stop the ventilator saving
selected parameters.
Measured Parameters Peak
Flow Airway
Pressure: Peak, Plateau, Mean, Base.
Expiratory Time.
Inspiratory Time and I:E Ratio of spontaneous
breaths. Expired
Tidal Volume. Expired
Minute Volume.
Graphic Display
Screen
Selection of pressure, volume and flow vs. time waveforms, and
pressure/volume and flow/volume loops.
|
Sweep speed
selection of horizontal axis to 3, 6 and 12 seconds. Three
scales for change the vertical axis. Timed graphics can be
frozen and all the screen printed to a PC file or compatible
printer. Trends of tidal volume, minute volume, rate,
peak and base pressure, peak flow and dynamic compliance of
the last 3, 6, 12 and 24 hours. Pressure bar graph
showing airway pressure with two scales (0-60, 0-120 cm H 2
O).
Alarms High and
Low Inspiratory Pressure High and
Low Tidal Volume High and
Low Minute Volume
Apnea High
breathing rate Low
PEEP Low
driving gas pressure Power
loss Battery
low charge
Technical Alarm
Silence: 3060 s The alarms
are activated according to levels of priorities with audible and visible (light) signals, and simultaneous written
messages on the screen.
Each event is documented in a list with date and hour.
Respiratory Mechanics AutoPEEP Static and
Dynamic Compliance Static and
Dynamic Resistance Slow Vital
Capacity P0.1
Electrical Voltage
range: 100 240 ACV (auto-switching) Frequency:
50-60 Hz; Power: 50 watts 12 V
internal rechargeable battery: Operating time: At least 30
minutes.
Pneumatic Pressurized Oxygen: 2.9 - 6.9 bar (42-100 psi). DISS
connector inlet. Pressurized Air: 2.9 - 6.9 bar (42-100 psi). DISS
connector inlet.
Standard Accessories Reusable
adult breathing circuit ( 22 mm) with water
traps, exhalation
flow sensor with expiratory valve. Flex arm
for breathing circuit. Air and
oxygen high pressure hose with DISS
connectors.
Compressed air water trap. Test
lung. Power
cord.
Pedestal: Four-wheel stand.
Operating instruction manual
Optional Accessories Heated
humidifier F&P 410.
Pediatric circuit ( 15 mm). Neonate circuit ( 12 mm).
Oilless compressor
FINI MED150
Warranty One year
warranty on material and workmanship
Price in Moscow
$ 19.970,-
|