         Neumovent Bebe Infant

 

     The NEUMOVENT Bebe Infant Ventilator is a microprocessor-controlled, pneumatically-operated, time-cycled, pressure-limited, continuous flow system with sigh capability device, which has been developed for use with infants, including premature newborns.

     The microprocessor conducts the activity of each electronic command and provides a comprehensive warnings and system error messages. The ventilatory parameters are preset through an independents election. The introduced data is shown in digital displays.

     During all modes of operation, the clinician can limit, safely, the delivered pressure. The O2 concentration is supply through an integral precision air-oxygen blender. Three power source with automatic switching, assures a safe ventilatory procedure.

 




   Continuous Mandatory Ventilation (CMV) Controlled, with Zero End Expiratory Pressure (ZEEP) or with Positive End Expiratory.

   Pressure (PEEP):
Ventilation with Intermittent. Positive Pressure and a breath ingrate from 1 to 150 breaths per minute .

   Patient-Triggered Assisted Ventilation: Triggering is accomplished by pressure.

   Synchronized Intermittent Mandatory Ventilation (SIMV) with or without Positive End Expiratory Pressure: Mechanical breath that can be given in between spontaneous breathing synchronized with the patient's demand.

   Continuous Positive Airway Pressure (CPAP): Spontaneous ventilation with positive pressure throughout the whole respiratory cycle, with or without backup ventilation programming .

   Manual Ventilation: Starts an inspiratory phase.

 

     Controls
Inspiratory Time:
From 0 .10 to 3 s, with Expiratory Time monitoring.
Rate:
Up to 150 breaths per minute.
I:E Ratio:
From :20 to 5:1
O. Percentage:
Continuous regulation from 21% to 100% (build-in a ir-oxygen m ixer). Precision: ±2%.
Inspiratory Flow:
Precision flowmeter: 0-15 L/min.

Inspiratory Pressure Limit: Continuous regulation from 5 to 80 cm H2O.
PEEP/CPAP:
Continuous regulation up to +25 cm H2O
ZEEP:
Continuous regulation up to -2 cm H2O
IMV/CPAP Flow:
Continuous regulation up to 30 L /min
Sigh:
1.5 x the programmed Ti; 5 times per hour.


     Alarms 
Low Pressure Alarm:
From 1 to 60 cm H2O
High Pressure Alarm:
From 5 to 80 cm H2O
Loss of PEEP/CPAP:
From 0 to 60 cm H2O
Oxygen Low Pressure:
Below 3 kg/cm
Air Low Pressure:
Below 3 kg/cm
Power Failure
Battery Low Charge
Micro Failure

     Glas Supply Pressure
Oxygen:
From 3.5 kg/cm to 7 kg/cm
Air :
3.5 kg/cm to 7 kg/cm

 

     Power Source
220-110 VAC
(internal bridge): Panel light ON when the plug is connected to mains.
Internal battery: Built-in battery 12 VDC with automatic recharging while the equipment is connected to mains.
External battery: Not included

Serial Communication
Type:
RS-232-C
Connector:
DB-9 (female).

 

     Accesories
Flowmeter 0-15 L/min.
Breathing circuit with wye, pressure tubes, tw Water traps and expiratory valve.o Water traps and expiratory valve. Test lung. High pressure hoses with connectors for oxygen and Air. Air source water trap inlet.
Four wheel stand.
Operating instruction manual.



     Optional
Heater-humidifier F&P 410

Oilless compressor FINI MED102



     Limited Warranty
Materials and workmanship for one year

 

     Dimensions and Weight
Heigh
(with stand 43" (109 cm)
Gabinet: 13½"(w) x 12½ "(d) x 9"(h) (35 x 32 x 23 cm)
Total weight (approx): Net: 48 lbs (22 kg); Gross: 67 lbs (30 kg).

 

     Price in Moscow     $ 9.900,-



Hosted by uCoz